Overview
- 3 Channel Controller for one ionization and two convection gauges or one ionization, one convection and one alternate capacitance diaphragm gauge
- Various pressure measurement ranges from 2.00E-11 to 1,000 Torr (2.7 x 10-11 to 1333 mbar)
- 3 analog outputs, 6 setpoint relays, RS232/RS485 serial communications, remote Digital I/O
- Compact space saving half rack design, bench top or panel/rack mount instrument
- Outstanding product warranty of 5 years
The Kurt Lesker Company (KJLC) series 3000 multi gauge controller is a highly versatile vacuum measurement controller capable of operating one ionization gauge (IG) and up to two KJL275 convection vacuum gauge sensors (CG), or one IG, one CG and one alternate gauge such as a capacitance diaphragm gauge (CDG). The ionization gauge on/off, degas functions and emission current selection can be controlled via front panel soft-keys, remote input signals (Digital I/O), or serial communications. IG sensor can also be automatically turned on/off using the measurements from one of the user selectable convection or alternate gauges. The MGC can be set to automatically switch (auto-ranging) between various emission currents. This results in optimal and stable pressure readings over the entire measurement range from low to high vacuum. High efficiency power supply design and effective thermal management techniques are used to enable operation of the MGC without the need for air movement devices such as troublesome fans. The state of all setpoint relays, emission current, filament in use and error messages for all fault conditions are displayed on the easy to read OLED set-up screen. Filament operation including filament current, emission current and ion current can be monitored in real time in the research screen mode.
MGC Controller | Ionization Gauge Compatibility |
Convection Gauge Compatibility |
MGC-3400 | One Nude UHV EB-degas Hot Cathode (2.00E-11 to 1.00E-03 Torr) (2.7 x 10-11 to 1.3 x 10-3 mbar) |
Up to two KJL275 Convection Gauges (1.00E-04 to 1000 Torr) (1.0 x 10-4 to 1,333 mbar) |
MGC-3300 |
One Nude I2R Hot Cathode (4.00E-10 to 1.00E-03 Torr) (5.3 x 10-10 to 1.3 x 10-3 mbar) or One Glass I2R Hot Cathode (4.00E-10 to 1.00E-03 Torr) (5.3 x 10-10 to 1.3 x 10-3 mbar) |
Up to two KJL275 Convection Gauges (1.00E-04 to 1000 Torr) (1.0 x 10-4 to 1,333 mbar) |
MGC-3200 | One KJLC351 Hot Cathode (1.0 x 10-9 to 5 x 10-2 Torr) (1.3 x 10-9 to 6.7 x 10-2 mbar) |
Up to two KJL275 Convection Gauges (1.00E-04 to 1000 Torr) (1.0 x 10-4 to 1,333 mbar) |
Specifications Table
Measurement Range: (Vacuum Gauge Dependent) |
MGC-3400: 2.00E-11 to 1,000 Torr (2.7 x 10-11 to 1333 mbar) when used with a UHV Nude EB-degas hot cathode IG and KJL275 CG MGC-3300: 4.00E-10 to 1,000 Torr (5.3 x 10-10 to 1333 mbar) when used with a Nude/Glass I2R hot cathode IG and KJL275 CG MGC-3200: 1.00E-09 to 1,000 Torr (1.3 x 10-9 to 1333 mbar) when used with the KJLC351 hot cathode IG and KJL275 CG |
Display - Pressure Indication Programming & Set-up Screen |
LED - 3 independent pressure display channels - 3 digit plus 2 digit exponent per channel OLED - displays state of all setpoint relays, IG emission current, error messages for fault conditions |
Units of Measure | Torr, mbar, Pa - user selectable |
Functionality |
IG: Powers & operates one KJLC Nude, Glass or KJLC351 ionization vacuum gauge, MGC-3400/3300 are also capable of operating other equivalent brands of nude/glass B-A hot cathode IG CG: Powers & operates up to 2 KJL275 convection or Granville-Phillips® (GP) Convectron® gauges Alternate Gauge: Displays pressure from an alternate gauge such as a CDG or other KJLC modules |
IG Sensor Control | IG sensor on/off, degas on/off and emission current can all be controlled via front panel soft-keys, remote input signals (digital I/O) or serial communications. IG sensor can also be automatically turned on/off using the measurements from one of the user selectable convection or alternate gauges. |
IG Remote Input Signals (Digital I/O) |
IG sensor on/off, degas on/off and emission current can also be set by applying momentary continuity to ground. Also the 9-pin D-sub remote input DIGITAL I/O connector provides pin-pin compatible signals with the GP 358 controller as well as compatible signals with the GP 307. |
Setpoint Relays/Contact Rating | Six single-pole, double-throw (SPDT) / 5 A at 30 Vdc, 5 A at 250 Vac, resistive load, user assignable to any of the gauges (Note - Contact rating applies to units shipped after Nov 1, 2015, See User Manual for older units). |
Analog Output |
Three analog outputs, user assignable to any of the gauges: IG Analog Output: Log linear 0 to 10 Vdc, 1 V/decade, various scaling selections also provide analog output compatibility with Granville-Phillips® controller models 307, 350 and 358 controllers, or log Linear 1.7 V to 9.3 Vdc (nominal 1.8 to 8.7 Vdc) 0.8 V/decade, or linear 0 to 10 Vdc (useable over 3 decades, also compatible with GP 307 controller) Combination IG + CG Analog Output: Log linear 0.5 to 7 Vdc, 0.5 V/decade CG Analog Output: Log linear 1 to 8 Vdc, 1 V/decade, or 0 to 7 Vdc, 1 V/decade (also compatible with GP 307, 350, 358), or non-linear analog S-curve 0.375 to 5.659 Vdc, or linear 0 to 10 Vdc (useable over 3 decades) |
Analog Input CDG KJLC Modules |
One 0-10 Vdc analog input signal from a CDG when used as an alternate gauge to CG2, or Analog input signal from one of the following KJLC module series: 275i, 300, 354 |
Serial Communications | RS485 / RS232 - ASCII protocol (command protocol compatibility with GP 307 and 358 controller is also provided) |
Status Output |
IG sensor on/off status message is displayed on the front panel, by serial communications and by SPDT relay (DIGITAL I/O Connector) rated at 1 A at 30 Vdc resistive, or 1 A at 30 Vac non-inductive. IG degas on/off status or IG error conditions are displayed on the front panel, by serial communications and by an open collector transistor output (ground emitter) rated at 40 V max. VCE, 50 mA IC max. |
IG Hot Cathode Filament Switching | User selectable between filament 1 or 2 using the front panel soft-keys or RS232/485 serial communications |
IG Hot Cathode Emission Current |
MGC-3400/3300: 100 µA, 4 mA, 10 mA or automatic switching (auto ranging) between 100 µA, 4 mA, 10 mA MGC-3200: 100 µA, 4 mA, or automatic switching (auto ranging) between 100 µA and 4 mA |
IG Degas |
MGC-3400: 40 W, electron bombardment MGC-3300: 40 W, resistive (I2R) MGC-3200: 3 W, electron bombardment |
IG Overpressure Protection |
MGC-3400/3300: Turns off nude/glass ion gauge filament at the following default settings: 1.00E-03 Torr at 100 µA emission current, 5.00E-04 Torr at 4 mA emission current, 1.00E-04 Torr at 10 mA emission current MGC-3200: Turns off KJLC351 ion gauge filament at the following factory default setting: 5.00E-02 Torr at 100 µA emission current and 1.00E-03 Torr at 4 mA emission current |
Temperature | Operating; 0 to + 40°C storage; -40 to + 70°C |
Humidity | 0 to 95% relative humidity, non-condensing |
Weight | 1.7 lb. (0.7 kg) |
Housing | Aluminum extrusion - black powder paint finish |
Input Power |
MGC-3400/3300: 20-28 Vdc, 200 W protected against power reversal and transient over-voltages MGC-3200: 20-28 Vdc, 36 W protected against power reversal and transient over-voltages |
Connectors |
IG & CG: Gauge cable assemblies provided by KJLC Digital I/O: 9-pin D-sub male Serial Communications: RS232; 9-pin D-sub female, RS485; 9-pin D-sub male Analog I/O, Setpoint Relays, Power: Pluggable terminal block (mating connectors included) |
Compliance & Standards |
CE compliant: EMC Directive 2004/108/EC, EN 61326-1, EN 55011 Low Voltage Directive 2006/95/EC, EN 61010-1 RoHS compliant |
351 Series Hot Cathode Bayard-Alpert Miniature-Ionization Vacuum Gauge
Overview
When used with KJLC's multigauge controllers, the 351 ionization vacuum gauge module provides the basic signal conditioning required to turn the gauge into a complete measuring instrument.
KJLC continuously strives to enhance the design, reduce cost and improve the performance and reliability of the 351. The electrometer circuit auto zeroes to ensure that the readings are not subject to temperature drift, eliminating the need for unnecessary, expensive circuitry which further reduces the cost.
All operations including filament on/off, degas, and change of emission current is controlled from the MGC.
The emission current can be set to automatically switch between 4 mA and 100 µA. This results in optimal and stabile pressure readings over the 351's measurement range, 1.0 x 10-9 to 5 x 10-2 Torr
The 351 gauge tube assembly is constructed of a compact all metal design with dual yttria coated Iridium or optional dual tungsten filaments.
For general vacuum applications, dual yttria coated filaments are offered for use with air and inert gases such as N2, argon, etc. Dual tungsten filaments are available for use with gases that may not be compatible with yttria coated filaments.
The gauge sensor assembly can be easily replaced in the field.
-
Wide measurement range
1.0 x 10-9 to 5 x 10-2 Torr
1.3 x 10-9 to 6.7 x 10-2 mbar
1.3 x 10-7 to 6.7 Pa - Ionization gauge designed specifically for use with KJLC's Multi-gauge Controller (MGC)
- Significant cost savings compared to other vacuum gauge measurement systems
- Full range measurement from 1 x 10-9 Torr to atmosphere when used with the KJLC Multi-gauge Controller (MGC) and a convection gauge
- Dual hot filament design, rugged and compact metal construction
- Built-in Electrometer results in significant controller and cabling cost reduction
Specifications Table
Measurement Range |
1.0 x 10-9 to 5 x 10-2 Torr / 1.3 x 10-9 to 6.7 x 10-2 mbar / 1.3 x 10-7 to 6.7 Pa |
Accuracy – N2 (Typical) | 1.0 x 10-8 to 5 x 10-2 Torr; ±15% of reading |
Repeatability - (Typical) | ±5% of reading |
Materials Exposed To Gases |
Dual Filaments: yttria coated iridium or optional tungsten Ion collector: tungsten Grid: tantalum Others: 316/304 SS, glass, nickel |
Sensitivity | Factory pre-set. Also user adjustable from 2 to 99 (set by the KJLC Multi-gauge Controller (MGC)) |
X-Ray Limit | < 5 x 10-10 Torr, < 6.7 x 10-10 mbar, < 6.7 x 10-8 Pa |
Emission Current | 100 µA, 4 mA, or automatic switching between 100 µA and 4 mA |
Degas | 3 W, electron bombardment |
Overpressure Protection |
Gauge is auto turned off at factory default setting of 5 x 10-2 Torr The MGC can also be set up to auto filament turn on/off using a convection gauge |
Filament Status | Filament on/off status is determined by LED on the 351 and also displays messages and available user interface options on the MGC |
Internal Gauge Volume | 1.0 in3 (16.4 cm3) |
Temperature | Operating: 0 to + 40°C; Storage: -40 to + 70°C |
Bakeout Temperature | 200°C (sensor only - electronics removed) |
Humidity | 0 to 95% relative humidity, non-condensing |
Weight | 0.6 lb. (0.27 kg) with NW25 KF flange |
Housing (Electronics) | Aluminum extrusion |
Mounting Orientation | Any |
Setpoint Relay | Relays available from MGC |
Input Signal | All 351 operations controlled from the MGC |
Filament Selection | User selectable between filament 1 and 2 - command signal sent from the MGC |
Input Power | Powered by MGC |
Connector/Cabling | KJLC cable/connector assembly for connection to MGC |
RF/EMI Protection | CE compliant |
Environmental | RoHS compliant |
275 Series Gauge Tubes
These gauge tubes and controllers are ideal for new applications or as low-cost drop-in replacements to the Convectron® brand of Granville-Phillips® gauges.
Features:
- Wide measuring range: 10-4 Torr to 1,000 Torr (10-4 to 1,333 mbar)
- Performance identical to existing competitor's gauges
- Less prone to damage from mechanical shock or vibration than competitor's tubes
- Less outgassing and faster response to pressure changes than competitor's tubes
- Rugged design withstands more abuse and harsh environments
- Significant savings—no changes to your system, you use your existing competitor's controllers or cables
- No changes to your process—a single vacuum gauge can monitor your vacuum system pumpdown and venting; helps keep your process up and running, even if it's not a laboratory environment
- CE marked, RoHS compliant
Specifications Table
Measurement Method | Convection Enhanced Pirani |
Measurement Range | 1 x 10-4 to 1,000 Torr (1.3 x 10-4 to 1,333 mbar) |
Temperature (operating) | 0-50° C |
Temperature (storage) | -40-70° C |
Temperature (bakeout) | 150° C maximum (gauge only, non-operating, electronics, and cable removed) |
Materials Exposed to Vacuum | Gold-plated tungsten, 304 & 316 stainless steel, glass, nickel, PTFE |
Accuracy (typical) |
1.0 x 10-4 to 1.0 x 10-3 Torr; 0.1 mTorr resolution 1.0 x 10-3 to 400 Torr; ± 10% of reading 400 to 1,000 Torr; ± 2.5% of reading |
Repeatability (Typical) | ± 2% of reading |
Mounting/Installation Orientation | Horizontal recommended |
Internal Volume | 26 cm3 (1.589 in3) |
Resolution | 1 x 10-4 |
Ion (Hot Filament) Gauge Tubes
Overview
Bayard-Alpert ion gauges have a reasonably linear response from 1 x 10-4 Torr to 1 x 10-9 Torr, with gauge sensitivities from 5 to 20 Torr-1. B-A gauges are available with one or two filaments (the second acting as a spare) and with two filament materials thoria-coated iridium, used in oxygen-rich applications and for "burn-out" protection if accidentally vented and tungsten, used for lower cost and in residual gases containing halogens.
Ordering Table
Description | Manual |
|
Part Number | Price | In Stock | Add To Cart | Accessories | ||||
---|---|---|---|---|---|---|---|---|---|---|---|
Description | Manual |
|
Part Number | Price | In Stock | Add To Cart | Accessories | ||||
Three channel pressure display vacuum gauge controller |
|
|
$1,744.00 | ||||||||
Three channel pressure display vacuum gauge controller |
|
|
$2,854.00 | ||||||||
Three channel pressure display vacuum gauge controller |
|
|
$2,888.00 |
Accessories Table
Description | Accessory Type | Part Number | Price | In Stock | Add Item | |
---|---|---|---|---|---|---|
Description | Accessory Type | Part Number | Price | In Stock | Add Item | |
Rack Mount Adapter panel for installation of one MGC as a left-mount or right-mount in a 2U, 19 inch wide rack |
|
|||||
Rack Mount Adapter panel for installation of two MGC controllers side-by-side in a 2U, 19 inch wide rack |
|
|||||
KJLC 351 to MGC, 10ft. (3m) |
|
|||||
KJLC 351 to MGC, 25ft. (8m) |
|
|||||
KJLC 351 to MGC, 50ft. (15m) |
|
|||||
KJLC 275 to Controller, 10ft. (3m) |
|
|||||
KJLC 275 to Controller, 25ft. (8m) |
|
|||||
KJLC 275 to Controller, 50ft. (15m) |
|
|||||
Glass Ion Gauge, DN35CF-DN40CF (2.75" CF), Single Iridium Filament |
|
|||||
Glass Ion Gauge, DN35CF-DN40CF (2.75" CF), Dual Tungsten Filaments |
|
|||||
Nude UHV Ion Gauge, DN35CF-DN40CF (2.75" CF), Dual Iridium Filaments |
|
|||||
1/4 Hard OFHC Gasket, DN16CF |
|
|||||
1/4 Hard OHFC Gasket, DN40CF |
|
|||||
Bolt & Nut Set for DN35CF-N40CF, 25 per package |
|
|||||
KJLC 275 Series Gauge Tube, DN16 ISO-KF - other flange types available |
|
|||||
KJLC 275 Series Gauge Tube, DN35CF-DN40CF (2.75" CF) - other flange types available |
|
|||||
KJLC 351 Ion Gauge, DN25 ISO-KF |
|
|||||
KJLC 351 Ion Gauge, DN35CF-DN40CF (2.75" CF) |
|
|||||
Power Supply for MGC-3200 - 110 AC cord included |
|
|||||
Power Supply for MGC-3200 - Universal, no AC power cord included |
|
|||||
Power Supply for MGC-3300 and 3400 -North American AC Plug included |
|
|||||
Power Supply for MGC-3300 and 3400 -Unlisted AC Plug Required |
|
|||||
Aluminum Clamp, DN16 ISO-KF Flange |
|
|||||
Viton® Centering Ring for DN16 ISO-KF Flange |
|
|||||
Aluminum Clamp, DN25 ISO-KF Flange |
|
|||||
Viton® Centering Ring for DN25 ISO-KF Flange |
|
|||||
Aluminum Clamp, DN40 ISO-KF Flange |
|
|||||
Viton® Centering Ring for DN40 ISO-KF Flange |
|
|||||
Bolt & Nut Set for DN16CF, 25 per package |
|
|||||
Glass Ion Gauge to MGC, 10ft. (3m) |
|
|||||
Glass Ion Gauge to MGC, 25ft. (8m) |
|
|||||
Glass Ion Gauge to MGC, 50ft. (15m) |
|
|||||
Nude Ion Gauge to MGC, 10ft. (3m) |
|
|||||
Nude Ion Gauge to MGC, 25ft. (8m) |
|
|||||
Nude Ion Gauge to MGC, 50ft. (15m) |
|
|||||
Nude Ion Gauge to MGC, 200C Bakeable, 10ft. (3m) |
|
|||||
Nude Ion Gauge to MGC, 200C Bakeable, 25ft. (8m) |
|
|||||
Nude Ion Gauge to MGC, 200C Bakeable, 50ft. (15m) |
|